User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Slaughter 104:AC Leakage Tester DATE: 08-Apr-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 13 NUMBER OF LINES: 152 CONFIGURATION: Datron 1281 STANDARD: Yokogawa 2262 STANDARD: Clarostat 240-C ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD Yokogawa 2262 1.004 STD Clarostat 240-C 1.005 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.006 DISP WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Set the UUT front panel controls as follows: 1.007 DISP POWER .............................................. OFF 1.007 DISP TEST VOLTAGE ................................. fully ccw 1.007 DISP CAPACITANCE COMPENSATION .......................... NONE 1.008 DISP Mechanically zero both UUT panel meters. 1.009 HEAD {} 1.010 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.011 HEAD {º VOLTAGE ACCURACY TEST º} 1.012 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.013 HEAD {}ÍÍ VOLTAGE ACCURACY TEST ÍÍ 1.014 JMP 2.001 1.015 EVAL 2.001 DISP ÿ Connectÿthe test equipment as follows: 2.001 DISP AC OUT HI (red) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]2262 6600V tap 2.001 DISP UUT GROUND (black) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]2262 V tap 2.001 DISP 2262 110V tap ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1281 HI 2.001 DISP 2262 v tapÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]1281 LO 2.002 DISP Set the UUT front panel controls as follows: 2.002 DISP POWER ............................................... ON 2.003 IEEE [@1281]ACV AUTO,FILT40HZ 2.004 DISP Adjust the TEST VOLTAGE control for a UUT reading 2.004 DISP of 2500 VAC. 2.005 ACC 2500.0V 0.2% 2.006 IEEE [@1281]RDG?[I] 2.007 MEM* 66 2.008 MEME 2.009 MEMC V 120.0U #! Test Tol 120, Sys Tol 5, TUR 24.000 (>= 4.00). 3.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 3.001 DISP of 2000 VAC. 3.002 ACC 2000.0V 0.2% 3.003 IEEE [@1281]RDG?[I] 3.004 MEM* 66 3.005 MEME 3.006 MEMC V 120.0U #! Test Tol 120, Sys Tol 4, TUR 30.000 (>= 4.00). 4.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 4.001 DISP of 1500 VAC. 4.002 ACC 1500.0V 0.2% 4.003 IEEE [@1281]RDG?[I] 4.004 MEM* 66 4.005 MEME 4.006 MEMC V 120.0U #! Test Tol 120, Sys Tol 3, TUR 40.000 (>= 4.00). 5.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 5.001 DISP of 1000 VAC. 5.002 ACC 1000.0V 0.2% 5.003 IEEE [@1281]RDG?[I] 5.004 MEM* 66 5.005 MEME 5.006 MEMC V 120.0U #! Test Tol 120, Sys Tol 2, TUR 60.000 (>= 4.00). 6.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 6.001 DISP of 500 VAC. 6.002 ACC 500.0V 0.2% 6.003 IEEE [@1281]RDG?[I] 6.004 MEM* 66 6.005 MEME 6.006 MEMC V 120.0U #! Test Tol 120, Sys Tol 1, TUR 120.000 (>= 4.00). 7.001 DISP Set the UUT front panel controls as follows: 7.001 DISP POWER .............................................. OFF 7.001 DISP TEST VOLTAGE ................................. fully ccw 7.002 DISP Disconnect the UUT from the Yokogawa 2262. 7.003 HEAD {} 7.004 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 7.005 HEAD {º MILLIAMMETER TEST º} 7.006 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 7.007 HEAD {}ÍÍ MILLIAMMETER TEST ÍÍ 7.008 JMP 8.001 7.009 EVAL 8.001 DISP ÿ Connectÿthe test equipment as follows: 8.001 DISP AC OUT HI (red) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ> 240-C HI 8.001 DISP ÿ 240-C LO ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ> 1281 I+ 8.001 DISP ÿ 1281 I- ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ> AC OUT LO (black) 8.002 DISP Set the 240-C decade switches to 0, 0, 0, 1, 0, 0. 8.003 DISP Set the UUT front panel controls as follows: 8.003 DISP POWER ............................................... ON 8.004 IEEE [@1281]ACI AUTO,FILT40HZ 8.005 DISP Adjust the TEST VOLTAGE control for a UUT reading 8.005 DISP of 5 mA AC. 8.006 ACC 20 5.000mA 300P% 100P/ 8.007 IEEE [@1281]RDG?[I] 8.008 MEM* 1000 8.009 MEME 8.010 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.0035, TUR 71.429 (>= 4.00). 9.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 9.001 DISP of 4 mA AC. 9.002 ACC 20 4.000mA 300P% 100P/ 9.003 IEEE [@1281]RDG?[I] 9.004 MEM* 1000 9.005 MEME 9.006 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.0032, TUR 78.125 (>= 4.00). 10.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 10.001 DISP of 3 mA AC. 10.002 ACC 20 3.000mA 300P% 100P/ 10.003 IEEE [@1281]RDG?[I] 10.004 MEM* 1000 10.005 MEME 10.006 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.0029, TUR 86.207 (>= 4.00). 11.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 11.001 DISP of 2 mA AC. 11.002 ACC 20 2.000mA 300P% 100P/ 11.003 IEEE [@1281]RDG?[I] 11.004 MEM* 1000 11.005 MEME 11.006 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.0026, TUR 96.154 (>= 4.00). 12.001 DISP Adjust the TEST VOLTAGE control for a UUT reading 12.001 DISP of 1 mA AC. 12.002 ACC 20 1.000mA 300P% 100P/ 12.003 IEEE [@1281]RDG?[I] 12.004 MEM* 1000 12.005 MEME 12.006 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.0023, TUR 108.696 (>= 4.00). 13.001 HEAD {} 13.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 13.002 DISP {º THIS COMPLETES THE VERIFICATION º} 13.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 13.003 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.